Which type of gas should i pick for a purified glove box?
The glove box should be supplied with inert gas with a 2 bars pressure. Normally, the inert gas chosen is nitrogen or argon and in some cases helium.
Regeneration requires a bottle composed either of hydrogen and inert gas or inert gas alone according to the glove box’ purification type;
Input |
Main feeding |
Reactor’s regeneration gas |
||
O2 Purification |
O2 and H2O Purification |
H2O Purification | ||
Gas |
Inert gas (Ar, N2,He) |
95 % Inert gas 5 % H2 |
100% Inert gas |
|
Pressure/Flow |
2 bars pressure |
Flow = 23 L.mn-1 |
||
Type of tank |
Tank or Inert gas network quality : [O2] < 5 ppm ; [H2O] < 5 ppm |
10 m3 Tank |
||
Connection |
Permanent |
Punctual (only during the reactor’s regeneration phase) |
||
Fitting |
4-6 or F1/4″ BSP flexible hose |
4-6 or F1/4″ BSP flexible hose |